compare specs

*   SS03-10   General purpose hydrophone ideal for kayaking, whale watching and full audio-band signal detection... more »
* SQ26-08 General purpose hydrophone ideal for kayaking, recreation, eco-art, and whale watching... more »
* CR1 Ideal for detection and recording of very loud signals, such as those from pile driving, air guns, or underwater demolition operations... more »
* CR2 Ideal for detection and recording of extremely loud signals, such as those from high-power sonar, large pile drivers, air gun arrays, or underwater explosions... more »
* C54XR
C54XRS
Cylindrical hydrophones for detection of audible and ultrasonic sounds, with the addition of an integral connector... more »
* C304 Directional receive pattern with very wide frequency response for monitoring ultrasonic signals not detectable by competing brands... more »

Hydrophone Systems:

*   Noise Measurement   calibrated underwater noise measurement systems - the only PC-based sound pressure level meter in existence to be in compliance with ANSI S1.11 and IEC1260 standards!... more »
* Underwater Recording high-quality underwater sound recording systems - from standard audio to ultrasonic frequencies... more »
* RUDAR™ Remote Underwater Digital Acoustic Recorder for underwater sound recording... more »
* Whale Watching portable listening systems for whale watching - an economical way to get set up quickly... more »

Cetacean Research Hydrophone Specifications Comparison



     
Model Number  
Linear Frequency Range [kHz] 0.0002 to 50 (±3dB)   0.002†† to 28 (±3dB)
Useable Frequency Range [kHz] 0.00002 to 110 (+3/-20dB)   0.0002†† to 250 (+3/-20dB)
Sensitivity [dB, re 1V/µPa] -199   -214
Power Requirement [Vdc] N/A   N/A
Maximum Operating Depth [m] 100   100 (preliminary)
Operating Temperature Range [ºC] -25 to 60†††   -40 to 70
Preamplifier Gain [dB] N/A   N/A
RMS Overload Acoustic Pressure [dB, re 1µPa] N/A   N/A
SPL Equiv. Self Noise at 1kHz [dB, re 1µPa] TBD   TBD
Capacitance [nF] or
Output Impedance [Ohms]
10.5nF   0.82nF
Dimensions [cm] 7.3L x 3.2 dia   5.6L x 1.4 dia
Coaxial Cable Length [m] 15
  10
Cable terminated with male BNC or mono phone plug connector
(longer cable optional)
Directionality
all are omnidirectional below 10kHz
-   -

Battery / Connector box

power not required
   

The CR1 requires a preamplifier with 100MOhms input impedance, such as the Reson VP1000 (PDF) or Reson VP2000 (PDF) preamplifiers. If a preamplifier with 1MOhms input impedance is used, then the low frequency -3dB point will be at 20Hz instead of 0.2Hz and the -20dB point will be at 2Hz instead of 0.02Hz.

††The CR2 also requires a preamplifier with 100MOhms input impedance, such as the Reson VP1000 (PDF) or Reson VP2000 (PDF) preamplifiers. If a preamplifier with 1MOhms input impedance is used, then the low frequency -3dB point will be at 200Hz instead of 2Hz and the -20dB point will be at 20Hz instead of 0.2Hz.

†††CR1 hydrophones are individually spot calibrated at the factory; calibration guaranteed between -5ºC and 30ºC.

CR2 hydrophones are not individually calibrated, but calibrated frequency response measurements can be performed upon request.









Model Number    
Linear Frequency Range [kHz] 0.016 to 50
(+2/-3dB)
122 to 250+
  0.016 to 44 (+2/-3dB)   0.007 to 250+ (±3dB)
Useable Frequency Range [kHz] 0.006 to 82
(+2/-20dB)
92 to 250+
  0.006 to 203 (+2/-20dB)   0.005 to 250+ (+3/-12dB)
Transducer Sensitivity* [dB, re 1V/µPa] -200   -185   -203
Preamplifier Gain [dB] 20   20
34
Effective Sensitivity* [dB, re 1V/µPa] -180   -165   -169
Power Requirement [Vdc] 5 to 32   5 to 32
5 to 32
Maximum Operating Depth ** [m] 460   460
460
Operating Temperature Range [ºC] -40 to 60   -40 to 60
-40 to 60
RMS Overload Acoustic Pressure [dB, re 1µPa] 184 to 201   169 to 186
173 to 190
SPL Equiv. Self Noise at 1kHz [dB, re 1µPa] 58   42 (< Sea State Zero)
~ 63
Capacitance [nF] or
Output Impedance [Ohms]
2.5Ohms   2.5Ohms
2.5Ohms
Dimensions [cm] 13.2L x (1.9 to 2.5) dia.   13.2L x (1.9 to 2.5) dia.
8.3L x 3.2W x (0.6 to 2.5)T
Shielded Cable Length [m] mating connector molded to 15, 30, or 50
(other lengths available on a custom order basis)

cable includes Integral Connector rated to a depth of 7,000 meters.

Directionality
all are omnidirectional below 10kHz

very directional at high frequencies (e.g., ~15º solid angle at 200 kHz, 3dB); directionality focused in a plane.

 

very directional at high frequencies (e.g., ~15º solid angle at 200 kHz, 3dB); directionality pattern is a dipole.

Battery / Connector box
rated NEMA-4X water resistant
(excluding connectors), requires
single 9V battery (not included)
9V battery life > 20 hrs

standard signal connection is through a 1/4" phone jack equipped with a water resistant cover or a watertight BNC connector-

external headphone amplifier available

 

standard signal connection is through a female BNC connector

Battery /connector box options integral water-resistent connectors
  high-pass filter switch built into battery box
  wire battery box to accept power from external supply
   

*  Transducer Sensitivity + Preamplifier Gain = Effective Sensitivity.

**  1,000 meter versions of the C54XR, C54XRS, and C304 hydrophones are available. Append "-1k" to the model number to order a hydrophone with a depth rating >1,000 meters.

C54 series and C304 Hydrophones are not individually calibrated, but calibrated frequency response measurements can be performed upon request.

 

T o o l s   f o r   R e s e a r c h ,   E d u c a t i o n    &   C o n s e r v a t i o n

home  |  hydrophones  |  analysis tools  |  accessories  |  suction cups  |  site map  |  top of page

Cetacean Research Technology
4728 12th Ave. NE - Seattle, WA 98105 USA

CONTACT CETACEAN RESEARCH TECHNOLOGY

privacy policy
copyright 1994, 1997-2008 Cetacean Research Technology